Error No. 7 - Maximum cycles are reached / Cause + Remedy?
Possible Cause:
The volume measurement is composed of
a plurality of individual cycles. The course of the measurement volume, the
maximum number of cycles has been reached, that a value could not be determined
for the volume.
!! Only for differential
pressure variant !!
Possible remedy:
-
Test volume is too large for
automatic volume measurement => Determine and enter volume manually
-
If possible, increase test pressure
for volume measurement => Change pressure parameter (observe max.
permissible pressure for test volume)